Publication | Open Access
Double Cycling During Mechanical Ventilation: Frequency, Mechanisms, and Physiologic Implications*
87
Citations
27
References
2018
Year
Double cycling is uncommon but occurs in all patients. Periods without double cycling alternate with periods with clusters of double cycling. The volume of the stacked breaths can double the set tidal volume in volume control-continuous mandatory ventilation with constant flow. Gas delivery must be tailored to neuroventilatory demand because interdependent ventilator setting-related physiologic factors can contribute to double cycling. One third of double-cycled breaths were reverse triggered, suggesting that repeated respiratory muscle activation after time-initiated ventilator breaths occurs more often than expected.