On-site calibration of the Raman laser absolute frequency for atom gravimeters

Yaoyao Xu, Jiafeng Cui, Kun Qi, Xiao-Bing Deng, Min-Kang Zhou, Xiao‐Chun Duan, Zhong-Kun Hu

Physical review. A/Physical review, A · 2018 · 14 citations · 33 references

Concepts

Abstract

The length standard for atom gravimeters is defined by Raman laser wavelength, which, in turn, depends on the absolute laser frequency. Here we present a method to measure Raman laser frequency based on an atom gravimeter itself, which has an advantage of calibrating the length standard on-site. The calibration utilizes the ${D}_{2}$ line of $^{87}\mathrm{Rb}$ atoms as a reference and takes advantage of the cold atom free fall as well as developed techniques of evaluating hyperfine-level shift within atom gravimeters. A calibration accuracy of 1 part in ${10}^{10}$ is achieved for the Raman laser frequency, which constrains the corresponding error below 1 $\ensuremath{\mu}\mathrm{Gal}$ in our gravity measurement.