Evaluation of ICP-OES Method for Heavy Metal and Metalloids Determination in Sterile Dump Material

D. Ilieva, Andriana Surleva, Manuela Murariu, Gabi Drochioiu, Mohd Mustafa Al Bakri Abdullah

Diffusion and defect data, solid state data. Part B, Solid state phenomena/Solid state phenomena · 2018 · 40 citations · 7 references

Concepts

Abstract

This study is aimed at development of an analytical method for the determination of heavy metals and arsenicin waste material from barite recovery by dual view inductively coupled plasma optical emission spectrometry (ICP-OES) after wet digestion. Aquaregia open digestion method was used for sample preparation. A soil certified reference material was used to study the efficiency of wet digestion procedure applied to sterile dump material. The sensitivity, accuracy, and precision of the method were determined. The achieved detection limits are: Arsenic (As) 0.80; Copper (Cu) 0.29; Lead (Pb) 0.29; Zink (Zn) 0.30; Cadmium (Cd) 0.11; Chromium (Cr) 0.09; Nickel (Ni) 0.02, mg.kg -1 .The developed method is applied for the determination of hazardous metals and arsenic in sterile dump material from closed barite mine in the region of Tarnita-Suceava, Romania.