Radiation Protection Dosimetry · 1993 · 13 citations · 0 references
Ec-wide SurveyNeonatologyRadiographyMedical ImagingMedicinePediatricsX Ray EquipmentPaediatric RadiologistsPaediatric RadiologySurgeryPediatric DosimetryRadiographic ImagingRadiographic Imaging CriteriaImage QualityRadiographic TechniqueNuclear MedicineRadiologyHealth Sciences
This EC-wide study surveyed the X ray equipment and the radiographic techniques used for seven frequent paediatric roentgen examinations in 89 departments of clinics and hospitals in 11 EC countries. Of the seven examinations investigated only skull AP/PA and chest AP/PA for a 10 month old infant will be reported here. Entrance surface dose (ESD) for these examinations was measured using CaF2 thermoluminescence dosemeters. The data were analysed based on recommended guidelines for good radiographic technique and image quality criteria as defined in the CEC Working Document 'Quality Criteria for Diagnostic Radiographic Images in Paediatrics'. A panel of eight paediatric radiologists rated the image quality of the original X ray films on the basis of above mentioned quality criteria. The analysis for all seven examinations showed that adherence to the guidelines on radiographic technique can lead to drastic and statistically significant reduction in ESD without any loss in image quality.