Radiation Protection Dosimetry · 1981 · 40 citations · 0 references
EngineeringMaterials ScienceNuclear EngineeringRadiation DetectionRadiation DosimetryRadiation MeasurementCharged Particle DetectionDetector SystemsNeutron DetectionPersonnel DosimetryProton DosimetryRadiation DetectorsChemical EtchingEtchingNeutron Detector Response FunctionNeutron Detector SystemsNeutron DetectorsNuclear RecoilRecoil LimitSensitivity
Commercially available etch track detectors offer significant potential for personnel neutron dosimetry. The monoenergetic neutron response of three - LR115, electrochemically etched CR-39, and polycarbonate - have been measured from 0.1 to 18 MeV. Significant differences in response are the result of differences in inherent charged particle sensitivity, track visualisation technique and pretreatment. The CR-39, made sensitive to protons by chemical pre-etching, has a threshold below 0.1 MeV. Polycarbonate, LR115 and CR-39 without pre-etching detect only heavy recoils (carbon, nitrogen and oxygen), and have thresholds at about 1 MeV. The response of LR115 is determined, to a large extent, by the degree to which a charged particle can penetrate the 12 µm detecting cellulose nitrate layer. Therefore, the sensitivity of LR115 increases almost exponentially with neutron energy (i.e. increasing recoil range). The responses of polycarbonate and CR-39 without pretreatment are much less sensitive to neutron energy above about 3 MeV. The inherent differences in these materials, together with CR-39 response changes induced by pretreatment parameter changes present new possibilities for personnel neutron spectrometry, as well as dosimetry.