Changes in the specific absorption rate (SAR) of radiofrequency energy in patients with retained cardiac leads during MRI at 1.5T and 3T

Laleh Golestanirad, Amir Ali Rahsepar, John E. Kirsch, Kenichiro Suwa, Jeremy D. Collins, Leonardo M. Angelone, Boris Keil, Rod Passman, Giorgio Bonmassar, Peter Serano,

Magnetic Resonance in Medicine · 2018 · 50 citations · 36 references

Abstract

For the models based on 6 patients studied, simulations suggest that MRI could be performed safely using a local head coil at both 1.5T and 3T, and with a body coil at 1.5T with pulses that produced <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mrow><mml:mrow><mml:mo>‖</mml:mo> <mml:mo>‖</mml:mo> <mml:mrow><mml:msubsup><mml:mi>B</mml:mi> <mml:mn>1</mml:mn> <mml:mo>+</mml:mo></mml:msubsup> </mml:mrow> <mml:mo>‖</mml:mo> <mml:mo>‖</mml:mo></mml:mrow> <mml:mo>≤</mml:mo> <mml:mn>4</mml:mn> <mml:mo> </mml:mo> <mml:mo>μ</mml:mo> <mml:mi>T</mml:mi></mml:mrow> </mml:math> . MRI at 3T could be performed safely in these patients using pulses with <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mrow><mml:mrow><mml:mo>‖</mml:mo> <mml:mo>‖</mml:mo> <mml:mrow><mml:msubsup><mml:mi>B</mml:mi> <mml:mn>1</mml:mn> <mml:mo>+</mml:mo></mml:msubsup> </mml:mrow> <mml:mo>‖</mml:mo> <mml:mo>‖</mml:mo></mml:mrow> <mml:mo>≤</mml:mo> <mml:mn>2</mml:mn> <mml:mo> </mml:mo> <mml:mo>μ</mml:mo> <mml:mi>T</mml:mi></mml:mrow> </mml:math> .

References

36