COMPUTER 3D-MODELING OF PATIENT-SPECIFIC NAVIGATIONAL TEMPLATE FOR CERVICAL SCREW INSERTION

A.V. Burtsev, Olga Pavlova, Sergey Ryabykh

Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika) · 2018 · 18 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

Objective. To prove the effectiveness of 3D computer modeling and printing of patient-specific templates for cervical spine screw fixation. Materials and Methods . Based on the MSCT data of three patients, 3D models of cervical vertebrae and guiding templates were produced. The screws were inserted transpedicularly into 3D-printed models of C 2 –C 7 vertebrae and C 1 lateral masses using standard cervical instrumentation. This technology was then clinically tested in a patient with C 2 tumor, in whom posterior stabilization of C 1 –C 3 vertebrae was performed using guiding templates. In the postoperative period, MSCT was performed to monitor the position of screws. Results. When implanting screws into 3D-models, technical difficulties were not revealed. After insertion of screws, a visual assessment of the 3D vertebra model was made. Malpositions in models were not detected. Based on the proven technique, a clinical approbation was performed in a patient with normal cervical spine pattern. MSCT study revealed the deviation of screws relative to the planned trajectory by no more than 2 mm, with no malpositions. Conclusion. The first domestic description of the technique of navigation using guiding templates is presented. Of all the existing methods, this is the most reliable way of screw positioning in the cervical spine. The possible insertion error does not affect the quality of positioning.

References