Publication | Open Access
Assessment of accuracy of out-of-field dose calculations by TiGRT treatment planning system in radiotherapy
31
Citations
15
References
2018
Year
This study highlights the limitations of TiGRT TPSs in calculating of the out-of-field dose. It should be noted that out-of-field data for this TPS should only be applied with a certain understanding of the accuracy of calculated dose outside the treatment field. Therefore, using the TPS-calculated dose could lead to an underestimation of secondary cancer risk as well as a weak clinical decision for patients with implantable cardiac pacemakers or pregnant patients.