Scientific Reports · 2018 · 49 citations · 42 references
EngineeringBiomedical EngineeringRadiologyBiomedical ImagingBiomedical OpticsDiagnostic ImagingMedical ImagingRadiographic ImagingX-ray OpticsDynamic ImagingIn Vivo ImagingPhase-contrast ImagingPhysiological MeasurementPhysiological StatusSmall Animal ImagingX-ray Imaging PhysicsX-ray Imaging Systems
We describe the first dynamic and the first in vivo X-ray imaging studies successfully performed at a laser-undulator-based compact synchrotron light source. The X-ray properties of this source enable time-sequence propagation-based X-ray phase-contrast imaging. We focus here on non-invasive imaging for respiratory treatment development and physiological understanding. In small animals, we capture the regional delivery of respiratory treatment, and two measures of respiratory health that can reveal the effectiveness of a treatment; lung motion and mucociliary clearance. The results demonstrate the ability of this set-up to perform laboratory-based dynamic imaging, specifically in small animal models, and with the possibility of longitudinal studies.
42
Phase-contrast imaging using polychromatic hard X-rays
S. Wilkins, Timur E. Gureyev, D. Gao et al. · Nature · 1996 · 1.8K citations
X-ray phase imaging with a grating interferometer
Timm Weitkamp, Ana Díaz, Christian Dávid et al. · Optics Express · 2005 · 1.2K citations · Full text
Demonstration of X-Ray Talbot Interferometry
Atsushi Momose, Shinya Kawamoto, Ichiro Koyama et al. · Japanese Journal of Applied Physics · 2003 · 964 citations
Natural Sciences, Materials Science, Diagnostic Radiology +16