Publication | Open Access
Development and Validation of RAPID: A Patient-Specific Monte Carlo Three-Dimensional Internal Dosimetry Platform
65
Citations
51
References
2018
Year
This work describes the development and validation of a patient-specific Monte Carlo internal dosimetry platform called RAPID (Radiopharmaceutical Assessment Platform for Internal Dosimetry). RAPID utilizes serial PET/CT or SPECT/CT images to calculate voxelized three-dimensional (3D) internal dose distributions with the Monte Carlo code Geant4. RAPID's dosimetry calculations were benchmarked against previously published S-values and specific absorbed fractions (SAFs) calculated for monoenergetic photon and electron sources within the Zubal phantom and for S-values calculated for a variety of radionuclides within spherical tumor phantoms with sizes ranging from 1 to 1000 g. The majority of the S-values and SAFs calculated in the Zubal Phantom were within 5% of the previously published values with the exception of a few 10 keV photon SAFs that agreed within 10%, and one value within 16%. The S-values calculated in the spherical tumor phantoms agreed within 2% for <sup>177</sup>Lu, <sup>131</sup>I, <sup>125</sup>I, <sup>18</sup>F, and <sup>64</sup>Cu, within 3.5% for <sup>211</sup>At and <sup>213</sup>Bi, within 6.5% for <sup>153</sup>Sm, <sup>111</sup>In, <sup>89</sup>Zr, and <sup>223</sup>Ra, and within 9% for <sup>90</sup>Y, <sup>68</sup>Ga, and <sup>124</sup>I. In conclusion, RAPID is capable of calculating accurate internal dosimetry at the voxel-level for a wide variety of radionuclides and could be a useful tool for calculating patient-specific 3D dose distributions.
| Year | Citations | |
|---|---|---|
2003 | 23.8K | |
2005 | 1.4K | |
1994 | 1.1K | |
2016 | 989 | |
2009 | 856 | |
2000 | 798 | |
1989 | 690 | |
1984 | 450 | |
2004 | 307 | |
2010 | 208 |
Page 1
Page 1