Asian Journal of Pharmaceutical and Clinical Research · 2018 · 10 citations · 0 references
PharmacyPharmaceutical AnalysisPharmacologyChromatographic AnalysisDissolution TestingPharmaceutical TechnologyAnalytical ChromatographyChromatographic Analysis Of PharmaceuticalsChromatographyIn Vitro Dissolution TestingLiquid ChromatographyLiquid Dosage FormsAnalytical Liquid ChromatographyChromatographic Drug AnalysisHigh-performance Liquid ChromatographyHigh-performance Liquid Chromatography AnalysisHigh-performance Liquid Chromatography Method Development
Objective: A simple, rapid, and reproducible high-performance liquid chromatography method was developed for the simultaneous determination of amlodipine and valsartan in their combined dosage forms and for drug dissolution studies.Methods: A C18 column (Zorbax Eclipse ХDB-C18, 5 μm, 2.1 mm × 150 mm) and a mobile phase of water:acetonitrile:trifluoroacetic acid (55:45:0.1 v/v/v) mixture were used for separation and quantification. Analyses were run at a flow rate of 0.4 mL/min and at ambient temperature. The injection volume was 5 μL and the ultraviolet detector was set at 265 nm. The method was validated as per ICH guidelines.Results: Under these conditions, amlodipine and valsartan were eluted at 1.64 min and 4.08 min, respectively. Total run time was shorter than 7 min. The results were 99.6 ± 0.6 and 98.5 ± 0.8 for amlodipine and valsartan, respectively. Valsartan was released within 15 min (98.32%) and amlodipine was also released within 30 min (96.16%) both at a pH of 6.8.Conclusion: The developed method was applied successfully for quality control assay of amlodipine and valsartan in their combination drug product and in vitro dissolution studies.