RADIATION DOSE AND IMAGE QUALITY IN PEDIATRIC HEAD CT
13
Citations
20
References
2018
Year
Computed TomographyPaediatric RadiologyPediatric Head CtTreatment VerificationCt ScanRadiation Therapy PlanningPediatric DosimetryRadiation ImagingCtdivol ValuesRadiologyHealth SciencesMedical ImagingRadiologic ImagingMedical Image ComputingRadiographic ImagingBiomedical ImagingPediatricsMedicineImage Quality
Our goal was to define a pediatric head CT protocol able to provide images of diagnostic quality, using the least amount of radiation, in children <10 years of age, while using a filtered back projection reconstruction algorithm. Image quality of 119 pediatric head CTs was assessed using a 5-point scoring system. Exams with scores ≥2.5 were considered of sufficient diagnostic quality. The contrast-to-noise ratio (CNR) was also measured. For children <1 year and 1-9 years, all studies performed with CTDIvol ≥ 20.1 mGy (range: 9-46 mGy) and CTDIvol ≥ 27.5 mGy (range: 15-60 mGy) yielded images of diagnostic quality. All diagnostic studies had a minimum CNR of 1.4. These CTDIvol values represent a good balance between image quality and radiation burden. This information can be helpful in designing pediatric head CT protocols with further dose-reduction, namely, iterative reconstruction algorithms and automated exposure control.