Journal of the Brazilian Chemical Society · 2018 · 10 citations · 0 references
Natural SciencesFood ScienceToxicologyAnalytical ChemistryFood AnalysisMass Spectrometric AnalysisAnalytical ChromatographyChromatographic Food AnalysisChromatographyEnvironmental Analytical ChemistryInstrumental AnalysisLiquid ChromatographyArsenic In FoodColumn ChromatographyEnvironmental Arsenic ContaminationHigh-performance Liquid ChromatographyHigh-performance Liquid Chromatography ColumnsLiquid Chromatography-mass SpectrometrySalt Gradient ElutionSalt Gradient Formation
This paper presents arsenic speciation studies by high-performance liquid chromatography with inductively coupled plasma-mass spectrometer (HPLC-ICP-MS) using a chromatography column containing a strong anion exchange stationary phase (Agilent Bio SAX G3154A/101) for determination of arsenobetaine (AsB), dimethylarsinic acid (DMA), As III , monomethylarsonic acid (MMA), and As V . The best separation was achieved in 15 min, using a 4 min isocratic step with 1 mM (NH 4 ) 2 HPO 4 pH 10.2, followed by a 2 min salt gradient from 1 to 10 mM, which was kept for 4 min. Compared to the Food and Drug Administration (FDA) method for rice using the PRP-X-100 column, the Bio SAX method was quicker, provided at least equal separation of AsB from the next peak, and led to more symmetrical peaks. The linearity between 0.5 and 60 g L -1 was shown. The method was tested with 10 g L -1 fortified rice and fish samples extracted in ultrasound using methanol/water 1:1, and recoveries between 40 and 90% were obtained.