Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy

J. Wyatt, Stephen Hedley, Emily Johnstone, R. Speight, Charles Kelly, Ann Henry, Susan Short, Louise Murray, David Sebag‐Montefiore, H. McCallum

Physics and Imaging in Radiation Oncology · 2018 · 24 citations · 20 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. MATERIALS AND METHODS: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. RESULTS: of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%-93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%-93%). CONCLUSIONS: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow.

References

20