Concepedia

Publication | Closed Access

Pure-Tone Audiometry With Forward Pressure Level Calibration Leads to Clinically-Relevant Improvements in Test–Retest Reliability

17

Citations

35

References

2018

Year

Abstract

These results indicate that reliability is better with insert earphones, especially with in-the-ear FPL calibration, compared with a standard clinical audiometer with supra-aural headphones. However, in-the-ear SPL calibration should not be used due to its sensitivity to standing waves. The improvement in reliability is clinically meaningful, potentially allowing hearing-conservation programs to more confidently determine significant threshold shifts at 6 kHz-a key frequency for the early detection of noise-induced hearing loss.

References

YearCitations

Page 1