Comparison of 2 Anti-PLA2R Immunoassays for the Diagnosis of Primary Membranous Nephropathy

Weihao Li, Guo Yaping, Zhiping Zhang, Feifei Zhang, Xiaomei Liu, Xin Ji, Lixia Liu, Hong Wang

Laboratory Medicine · 2018 · 13 citations · 11 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: Anti-phospholipase A2 receptor (anti-PLA2R) is a promising biomarker for diagnosis, activity evaluation, therapy monitoring, and prognostic estimation of primary membranous neuropathy (pMN). Difference in methodology may be one cause of the discrepancy in anti-PLA2R-positive percentages in reported studies. In this study, we evaluated the determination consistency of anti-PLA2R using indirect immunofluorescence assay (IIF) and enzyme-linked immunosorbent assay (ELISA). METHODS: A total of 113 patients with pMN, 34 with secondary membranous neuropathy (sMN), and 53 healthy control individuals were enrolled. We tested their circulating anti-PLA2R, along with other biochemical parameters, using IIF and ELISA. RESULTS: The sensitivity of anti-PLA2R for pMN was 70.8% by IIF and 67.3% by ELISA, with no statistically significant difference. The overall qualitative agreement of anti-PLA2R was 91.15% (95% confidence interval [CI], 85.91%-96.39%), and the correlation coefficient was 0.79 for IIF versus ELISA. The overall correlation coefficient of anti-PLA2R titers by IIF and concentration by ELISA was 0.85. Through ROC curve analysis, anti-PLA2R as measured by IIF and ELISA showed larger areas under the curve (AUCs) than other biochemical parameters. CONCLUSION: ELISA shows similar performance to IIF, with the advantages of quantitative results and suitability for high throughput.