Electronic Cortisol Detection Using an Antibody-Embedded Polymer Coupled to a Field-Effect Transistor

Hyun‐June Jang, Taein Lee, Jian Song, Luisa M. Russell, Hui Li, Jennifer Dailey, Peter C. Searson, Howard E. Katz

ACS Applied Materials & Interfaces · 2018 · 81 citations · 23 references

Concepts

Abstract

A field-effect transistor-based cortisol sensor was demonstrated in physiological conditions. An antibody-embedded polymer on the remote gate was proposed to overcome the Debye length issue (λD). The sensing membrane was made by linking poly(styrene-co-methacrylic acid) (PSMA) with anticortisol before coating the modified polymer on the remote gate. The embedded receptor in the polymer showed sensitivity from 10 fg/mL to 10 ng/mL for cortisol and a limit of detection (LOD) of 1 pg/mL in 1× PBS where λD is 0.2 nm. A LOD of 1 ng/mL was shown in lightly buffered artificial sweat. Finally, a sandwich ELISA confirmed the antibody binding activity of antibody-embedded PSMA.

References