Quantitative Assessment of Immunohistochemistry Laboratory Performance by Measuring Analytic Response Curves and Limits of Detection

Seshi R. Sompuram, Kodela Vani, Anika K. Schaedle, Anuradha Balasubramanian, Steven A. Bogen

Archives of Pathology & Laboratory Medicine · 2018 · 32 citations · 29 references

DOIFull text

Open access

Concepts

TL;DR

Interlaboratory variability in diagnostic immunohistochemistry (IHC) testing remains high, and the lack of quantitative measures of immunostain sensitivity hampers standardization. This study aimed to quantitatively and objectively assess the sensitivity of HER2, ER, and PR immunostains across IHC laboratories using a proficiency‑testing format. Thirty‑nine laboratories stained three slides—one for each marker—each containing a positive tissue section and IHControls at five defined concentrations of microbeads, and sensitivity was evaluated via analytic response curves and limits of detection. The analytic response curves showed strong reproducibility within labs but revealed up to 100‑fold sensitivity differences between vendors, and quantitative assessment proved superior to subjective limits of detection, providing an objective tool for identifying outlier laboratories.

Abstract

- Numerous studies highlight interlaboratory performance variability in diagnostic immunohistochemistry (IHC) testing. Despite substantial improvements over the years, the inability to quantitatively and objectively assess immunostain sensitivity complicates interlaboratory standardization.- To quantitatively and objectively assess the sensitivity of the immunohistochemical stains for human epidermal growth factor receptor type 2 (HER2), estrogen receptor (ER), and progesterone receptor (PR) across IHC laboratories in a proficiency testing format. We measure sensitivity with parameters that are new to the field of diagnostic IHC: analytic response curves and limits of detection.- Thirty-nine diagnostic IHC laboratories stained a set of 3 slides, one each for HER2, ER, and PR. Each slide incorporated a positive tissue section and IHControls at 5 different concentrations. The IHControls comprise cell-sized clear microbeads coated with defined concentrations of analyte (HER2, ER, and/or PR). The laboratories identified the limits of detection and then mailed the slides for quantitative assessment.- Each commercial immunostain demonstrated a characteristic analytic response curve, reflecting strong reproducibility among IHC laboratories using the same automation and reagents prepared per current Good Manufacturing Practices. However, when comparing different commercial vendors (using different reagents), the data reveal up to 100-fold differences in analytic sensitivity. For proficiency testing purposes, quantitative assessment using analytic response curves was superior to subjective interpretation of limits of detection.- Assessment of IHC laboratory performance by quantitative measurement of analytic response curves is a powerful, objective tool for identifying outlier IHC laboratories. It uniquely evaluates immunostain performance across a range of defined analyte concentrations.

References

29