Comparison of Manual and Fully Automated AIX1000 Rapid Plasma Reagin Assays for Laboratory Diagnosis of Syphilis

Alan M. Sanfilippo, Kristie Freeman, John L. Schmitz

Journal of Clinical Microbiology · 2018 · 19 citations · 22 references

DOIFull text

Open access

Concepts

Abstract

= 0.03). Among the 20 concordant reactive specimens, 68.4% of the titers agreed within ±1 dilution between methods. Reproducibility testing of the AIX1000 system demonstrated qualitative and semiquantitative (within ±1 dilution) agreement between specimens tested on different days of 96.0% and 76.0%, respectively, and 100.0% agreement between replicates within the same run. One of 24 samples analyzed under other disease conditions was reactive on both the AIX1000 system and the ASI card. Overall, the fully automated AIX1000 system demonstrated significantly enhanced sensitivity and specificity similar to that of the manual ASI RPR card test, making the AIX1000 system suitable for the laboratory diagnosis of syphilis in a clinical laboratory setting.