MedicineGynecologyRadiation OncologyRadiation DosimetryRadiation ProtectionTreatment PlanningPatient Dose OptimizationPatient-specific DosimetryPlan OptimizationPlan Quality AssuranceRadiation Treatment PlanningTreatment Planning PhysicsPlan Quality AssessmentPlan Quality MetricsRadiotherapy Plan EvaluationTreatment Plan EvaluationTreatment Plan OptimizationTreatment Planning DosimetryVolumetric Modulated Arc Therapy DosimetryVolumetric Modulated Arc Therapy Planning
AIM: tool included in the Eclipse treatment planning system, while preserving the plan quality. BACKGROUND: The treatment planning system Eclipse (Varian Medical Systems, Palo Alto, CA) includes a control mechanism for the number of monitor units of volumetric modulated arc therapy (VMAT) plans, named the MU Objective tool. MATERIAL AND METHODS: -test and changes in plan quality were assessed with a set of parameters for OARs and PTVs. RESULTS: = 0.002). Although similar MU reductions were observed for both sites, for some gynecological plans maximum differences were greater than 10%. All the assessed parameters for PTVs and OARs sparing showed average differences below 2%. CONCLUSION: = 400 led to the optimum MU reduction, maintaining the original dose distribution and dosimetric parameters practically unaltered.