Concepedia

Publication | Open Access

<scp>ASVCP</scp> guidelines: Allowable total error hematology

89

Citations

21

References

2018

Year

Abstract

The purpose of this document is to provide total allowable error (TE<sub>a</sub> ) recommendations for commonly analyzed hematology measurands for veterinary personnel. These guidelines define relevant terminology and highlight considerations specific to hematology measurands. They also provide reasons and guidelines for using TE<sub>a</sub> in instrument performance evaluation, including recommendations for when the total observed error exceeds the recommended TE<sub>a</sub> . Biological variation-based quality specifications are briefly discussed. The appendix describes the derivation of the hematology TE<sub>a</sub> recommendations and provides resources for external quality assurance/proficiency testing programs and a worksheet for implementation of the guidelines.