Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method

Anand Chandrasekhar, Chang‐Sei Kim, Mohammed B. Naji, Keerthana Natarajan, Jin‐Oh Hahn, Ramakrishna Mukkamala

Science Translational Medicine · 2018 · 211 citations · 17 references

Concepts

Abstract

= 32). We likewise tested a finger cuff device, which uses the volume-clamp method of BP detection. About 90% of the users learned the finger actuation required by the smartphone-based device after one or two practice trials. The device yielded bias and precision errors of 3.3 and 8.8 mmHg for systolic BP and -5.6 and 7.7 mmHg for diastolic BP over a 40 to 50 mmHg range of BP. These errors were comparable to the finger cuff device. Cuff-less and calibration-free monitoring of systolic and diastolic BP may be feasible via a smartphone.