MedicineInfectious DiseasesSurgeryCardiothoracic SurgeryMedical MicrobiologyPerioperative Infection PreventionBacterial InfectionsImplant-associated InfectionImplant-related Infection PreventionInfective EndocarditisPostoperative Infection PreventionSurgical Infection ManagementInfection Risk FactorsIntracardiac Device-related EndocarditisPostoperative Infectious ComplicationsProsthetic InfectionSurgical Site Infection After ImplantationTransmission Risk FactorsWater ContaminationWater Source Contamination
from the exhaust air of intrinsically contaminated HCDs to patients during cardiac surgery. To date, all reported cases have been associated with Stöckert 3T HCDs manufactured at one plant by LivaNova prior to September 2014. Implantation of prosthetic material increases the risk of infection. Infections usually present as prosthetic valve endocarditis, vascular graft infection or disseminated infection. Reported mortality rates have varied, but were often over 40%. Several measures are recommended to facilitate case-finding and mitigate risk of exposure. The feasibility of some risk mitigation measures and their effectiveness in reducing the risk of exposure are yet to be determined. Until HCDs are redesigned in a manner that prevents water contamination and aerosolization, separating the HCD exhaust air from the operating room air during surgery may be the most effective risk mitigation strategy. However, possible unintended consequences of this approach should be considered. This overview summarizes findings from peer-reviewed and other relevant national documents on key features of the outbreak, including the source, identified risk factors for infection, signs and symptoms of infection, burden of disease, risk mitigation measures, management challenges and knowledge gaps.
15
Philipp Köhler, Stefan P. Kuster, Guido V. Bloemberg et al. · European Heart Journal · 2015 · 212 citations · Full text