Concepedia

Publication | Closed Access

The True Ablation Effect of Holmium YAG Laser on Soft Tissue

52

Citations

9

References

2018

Year

Abstract

The overall mean ID was significantly higher than the theoretical 0.4 mm PD described for Ho:YAG laser. The energy, frequency, and pulse length had individual effects regarding ID, incision width, and coagulation. The ID should be specified in accordance with the laser's power output and should not be confused with the physics of PD concept.

References

YearCitations

Page 1