Concepedia

Publication | Closed Access

21 Tesla FT-ICR Mass Spectrometer for Ultrahigh-Resolution Analysis of Complex Organic Mixtures

320

Citations

63

References

2018

Year

Abstract

We describe complex organic mixture analysis by 21 tesla (T) Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS). Ultrahigh mass-resolving power (m/Δm<sub>50%</sub> > 2 700 000 at m/z 400) and mass accuracy (80 ppb rms) enable resolution and confident identification of tens of thousands of unique elemental compositions. We demonstrate 2.2-fold higher mass-resolving power, 2.6-fold better mass measurement accuracy, and 1.3-fold more assigned molecular formulas compared to our custom-built, state-of-the-art 9.4 T FT-ICR mass spectrometer for petroleum and dissolved organic matter (DOM) analyses. Analysis of a heavy petroleum distillate exemplifies the need for ultrahigh-performance mass spectrometry (49 040 assigned molecular formulas for 21 T versus 29 012 for 9.4 T) and extends the identification of previously unresolved O<sub>o</sub>, S<sub>s</sub>O<sub>o</sub>, and NO<sub>o</sub> classes. Mass selective ion accumulation (20 Thompson isolation) of an asphalt volcano sample yields 462 resolved mass spectral peaks at m/z 677 and reveals previously unresolved C<sub>c</sub>H<sub>h</sub>N<sub>n</sub>O<sub>o</sub>S<sub>s</sub> mass differences at high mass (m/z > 600). Similar performance gains are realized in the analysis of dissolved organic matter, where doubly charged O<sub>o</sub> species are resolved from singly charged SO<sub>o</sub> species, which requires a mass-resolving power greater than 1 400 000 (at m/z 600). This direct comparison reveals the continued need for higher mass-resolving power and better mass accuracy for comprehensive molecular characterization of the most complex organic mixtures.

References

YearCitations

Page 1