Concepedia

Publication | Open Access

Deadtime effects in quantification of 177Lu activity for radionuclide therapy

24

Citations

16

References

2018

Year

Abstract

The paralyzable model showed to be appropriate for the investigated range of counts, which were five to six times higher than those observed in the patient post-therapy imaging. Our results suggest that the deadtime corrections should be based on count losses in the scatter-corrected photopeak window and not on the deadtime determined from the full spectrum. Finally, a general procedure that can be followed to correct patient images for deadtime is presented.

References

YearCitations

Page 1