Pneumatic low-coherence interferometry otoscope to quantify tympanic membrane mobility and middle ear pressure

Jungeun Won, Guillermo L. Monroy, Pin‐Chieh Huang, Roshan Dsouza, Malcolm C. Hill, Michael Novák, Ryan G. Porter, Eric J. Chaney, Ronit Barkalifa, Stephen A. Boppart

Biomedical Optics Express · 2018 · 16 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

test. As a result, the compliance represents pneumatic TM mobility, and the amplitude ratio estimates MEP. The presence of a middle ear effusion (MEE) significantly decreased compliance (p<0.001). The amplitude ratio of the OM group was statistically less than that of the normal group (p<0.01), indicating positive MEP. Unlike tympanometry, pneumatic LCI otoscopy quantifies TM mobility as well as MEP regardless of MEE presence. With combined benefits of pneumatic otoscopy and tympanometry, pneumatic LCI otoscopy may provide new quantitative metrics for understanding TM dynamics and diagnosing OM.