MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingJoint Replacement SurgeryOrthopedicsFracture BiomechanicsFracture FixationFracture ManagementFracture MechanicsRevision ArthroplastyRevision Joint ReplacementBone Fracture RiskFemur FracturesFracture AnalysisFracture Surface AnalysisFracturesImplant Failure ManagementModular Revision ImplantsRevision Risk
Fractures of well-ingrown femoral components are a rare and often challenging complication after revision total hip arthroplasty. Prior series have documented catastrophic failure at the modular junction of revision femoral components. However, to the authors' knowledge, there has been only 1 report of a mid-stem fracture of a modular tapered revision stem. The present article reports 2 cases of fatigue fractures (14 months and 10 years after implantation) of a tapered modular revision stem. It presents the results of the fracture surface analysis, discusses the etiology of failure, and presents the authors' recommendations on how to best avoid this complication.
14
Prosthetic component failures in hip arthroplasty surgery
David A. Heck, Cynthia M. Partridge, Jeffrey D. Reuben et al. · The Journal of Arthroplasty · 1995 · 161 citations