Blood Pressure Monitoring · 2017 · 36 citations · 5 references
MedicineCardiologyMetrologyNorm-referenced AssessmentPhysiatryBlood Pressure ControlBlood Pressure ManagementCardiovascular AssessmentCardiovascular MonitoringPressure MeasurementBlood PressureBlood Pressure Cuff ValidationBlood Pressure MeasurementBlood Pressure MonitoringBlood Pressure MonitorsBlood Pressure Normative Reference ValuesNoninvasive Blood Pressure MeasurementPediatric Blood Pressure Measurement
OBJECTIVE: To assess the accuracy of the professional oscillometric blood pressure (BP) monitor Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. METHODS: Adults and children (aged 3-12 years) were recruited to fulfil the age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard using the same-arm sequential BP measurement method. Three cuffs of the test device were used for arm circumferences of 14-22, 22-32 and 32-42 cm. RESULTS: A total of 115 participants were recruited and 88 were included in the analysis (51 adults and 37 children). For criterion 1, the mean±SD of the differences between the test device and reference BP was -1.0±7.0/-4.7±5.4 mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference per participant was 5.15/4.77 mmHg (systolic/diastolic). CONCLUSION: The professional Microlife WatchBP Office BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard for adults and children and can be recommended for clinical use.
5
George S. Stergiou, Dimitris Tzamouranis, Efthimia Nasothimiou et al. · Blood Pressure Monitoring · 2008 · 119 citations
George S. Stergiou, Che Lin, Chia-Ming Lin et al. · Blood Pressure Monitoring · 2008 · 20 citations