Journal of Spine Surgery · 2017 · 11 citations · 14 references
MedicineOrthopaedic SurgerySurgeryBone GraftingImplant SurgerySpine SurgeryCervical SurgerySpinal FusionAnterior Plate FixationCervical Disc FusionCervical Spinal FusionOperative TechniquePostoperative RadiographsRadiographic Follow-up AssessmentRadiographic ReviewSubsidence
BACKGROUND: Anterior cervical discectomy and fusion (ACDF) may be performed using an interbody cage or graft with an anterior plate or with a stand-alone (SA) interbody device without the anterior plate. The pros and cons of each vary. This study examined the radiographic outcome of the two techniques with a focus on implant subsidence. METHODS: A retrospective review of cases of singe level ACDF by a single surgeon was undertaken. Medical and radiographic records were reviewed to determine subsidence, pre- and post-operative segmental and total lordosis in cohorts of both stand-alone and graft-and-plate constructs. RESULTS: The post-operative radiographs of 35 patients with a SA cage were compared with 41 patients with an allograft block and anterior plate (graft and plate; GP). There was no significant difference in overall subsidence between the two groups although there was a trend toward less clinically significant subsidence (2 mm) in the SA group. For single level ACDF, a SA device appears to be comparable in terms of undesired subsidence. CONCLUSIONS: Further studies with different implants and materials may offer further insight.