Publication | Open Access
A comparison of two pencil beam scanning treatment planning systems for proton therapy
20
Citations
16
References
2017
Year
Both Eclipse and Raystation PB TPS are not always accurate within ±3% for a 5 cm range shifters or for small targets. This was improved with the Raystation MC model. The point dose calculations of Eclipse, Raystation PB, and Raystation MC compare within ±3% to measured doses for the other scenarios tested.
| Year | Citations | |
|---|---|---|
Page 1