Concepedia

Publication | Open Access

Patient-Perceived and Laboratory-Measured Halos Associated with Diffractive Bifocal and Trifocal Intraocular Lenses

52

Citations

16

References

2017

Year

Abstract

The addition power of the tested IOLs affected both laboratory-measured and patient-perceived halos. Trifocal lenses generated fewer complaints about halos. Precise: Multifocal intraocular lenses (MIOLs) give rise to halos observed by patients under conditions of dim lighting. This study compares laboratory characterized halos generated by five MIOLs with patient-perceived halos.