Accuracy of the raw-data-based effective atomic numbers and monochromatic CT numbers for contrast medium with a dual-energy CT technique
16
Citations
14
References
2017
Year
The accuracy of the raw-data-based Z<sub>eff</sub> values was higher than that of image-based Z<sub>eff</sub> values in the tissue-equivalent phantom. The accuracy of Z<sub>eff</sub> values in the contrast medium was in good agreement within the maximum SD found in the iodine concentration range of clinical dynamic CT imaging. Moreover, the optimum monochromatic energy for human tissue and iodinated contrast medium was found to be 70 keV. Advances in knowledge: The accuracy of the Z<sub>eff</sub> values and monochromatic CT numbers of the contrast medium created by raw-data-based, dual-energy CT could be sufficient in clinical conditions.