2017 IEEE International Ultrasonics Symposium (IUS) · 2017 · 11 citations · 10 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingUltrasonographyUltrasoundDiagnostic UltrasonographyMagnetic Resonance ImagingMedical ImagingMedical Imaging InstrumentationUltrasound ImagingUltrasound InstrumentationBandwidthFerromagnetic ResonanceFractional BandwidthMagnetic ResonanceMagnetic Resonance Imaging InstrumentationThree-dimensional Ultrasound ImagingTransducer Array DesignUltrasound Image Acquisition
We developed a magnetic resonance compatible real-time, three-dimensional imaging ultrasound probe for motion management of radiation therapy for liver cancer. The probe contains an 18,000-element, 46.8 mm × 21.5 mm matrix array constructed from three tiled transducer modules with integrated beamforming ASICs. The center frequency and -6 dB fractional bandwidth of the probe was 3.6 MHz and 85 percent respectively. Ferromagnetic materials in the acoustic stack, flex interconnect and electronics boards were greatly minimized for magnetic resonance compatibility. The probe and cable were shielded to minimize the impact of radiofrequency noise on both the ultrasound and magnetic resonance images. The probe's low-profile, side-viewing design allows it to be strapped to a patient so that images may be acquired hands-free. We present simultaneously acquired ultrasound and 3 Tesla magnetic resonance images with minimal artifacts in both images.
10
Stereotactic Radiotherapy of Malignancies in the Abdomen: Methodological aspects
Ingmar Lax, Henric Blomgren, Ingemar Näslund et al. · Acta Oncologica · 1994 · 546 citations · Full text
J J W Lagendijk, Bas W. Raaymakers, Alexander Raaijmakers et al. · Radiotherapy and Oncology · 2007 · 509 citations