Evaluation of an Integrated System of Wearable Physiological Sensors for Stress Monitoring in Working Environments by Using Biological Markers

Stefano Betti, Raffaele Molino Lova, Erika Rovini, Giorgia Acerbi, Luca Santarelli, Manuela Cabiati, Silvia Del Ry, Filippo Cavallo

IEEE Transactions on Biomedical Engineering · 2017 · 217 citations · 38 references

DOIFull text

Open access

Concepts

TL;DR

The study develops and tests a wearable ECG, EDA, and EEG sensor system to detect human stress and evaluates its correlation with salivary cortisol levels. Fifteen healthy adults performed the Maastricht Acute Stress Test while wearing three commercial sensors; salivary samples were collected throughout, and extracted physiological features were analyzed using support vector machine classification and correlation analysis. The sensor system achieved 86 % classification accuracy and a correlation coefficient of 0.714 with salivary cortisol, demonstrating reliable stress detection and potential for remote medical interventions.

Abstract

The objectives of this paper are to develop and test the ability of a wearable physiological sensors system, based on ECG, EDA, and EEG, to capture human stress and to assess whether the detected changes in physiological signals correlate with changes in salivary cortisol level, which is a reliable, objective biomarker of stress.15 healthy participants, eight males and seven females, mean age 40.8 ± 9.5 years, wore a set of three commercial sensors to record physiological signals during the Maastricht Acute Stress Test, an experimental protocol known to elicit robust physical and mental stress in humans. Salivary samples were collected throughout the different phases of the test. Statistical analysis was performed using a support vector machine (SVM) classification algorithm. A correlation analysis between extracted physiological features and salivary cortisol levels was also performed.15 features extracted from heart rate variability, electrodermal, and electroencephalography signals showed a high degree of significance in disentangling stress from a relaxed state. The classification algorithm, based on significant features, provided satisfactory outcomes with 86% accuracy. Furthermore, correlation analysis showed that the observed changes in physiological features were consistent with the trend of salivary cortisol levels (R2 = 0.714).The tested set of wearable sensors was able to successfully capture human stress and quantify stress level.The results of this pilot study may be useful in designing portable and remote control systems, such as medical devices used to turn on interventions and prevent stress consequences.

References

38