Concepedia

Publication | Open Access

An Evaluation of Portable Wet Bulb Globe Temperature Monitor Accuracy

46

Citations

10

References

2017

Year

Abstract

The QUESTemp° 34, followed by the Extech, had the smallest error compared with the reference unit. Moreover, the QUESTemp° 34, Extech, and Kestrel units appeared to offer conservative yet accurate assessments of the WBGT, potentially minimizing the risk of allowing physical activity to continue in stressful heat environments. Instruments using the psychrometric wet bulb temperature tended to underestimate WBGT under low wind-speed conditions. Accurate WBGT interpretations are important to enable clinicians to guide activities in hot and humid weather conditions.

References

YearCitations

Page 1