IEEE Sensors Letters · 2017 · 26 citations · 18 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsElectrochemistrySensorsBlood Glucose MonitoringContinuous Glucose MonitoringElectrochemical Impedance SpectroscopyElectrochemical SensorsGlycemic MonitoringImpedance SpectroscopyCharge Transfer ResistanceDouble-layer CapacitanceElectrical Double Layer CapacitanceElectrical ImpedanceGlucose MonitoringGlucose SensingGlucose SensorsSensor Sensitivity
In this article, we study the change in the sensitivity and the electrochemical impedance of continuous glucose monitoring sensors over time. In total, 28-day sensitivity and electrical impedance spectroscopy measurement results on four similar sensors are presented. The sensitivity of the sensor is observed to be related to its double-layer capacitance and charge-transfer resistance, based on results acquired from a sensor that showed substantial sensitivity drop. Two data clusters are extracted that relate the sensor sensitivity to its impedance before and after the sensitivity drops by more than 50%.