Study of Electrochemical Impedance of a Continuous Glucose Monitoring Sensor and its Correlation With Sensor Performance

Sara S. Ghoreishizadeh, Xiaotian Zhang, Sanjiv Sharma, Pantelis Georgiou

IEEE Sensors Letters · 2017 · 26 citations · 18 references

Concepts

Abstract

In this article, we study the change in the sensitivity and the electrochemical impedance of continuous glucose monitoring sensors over time. In total, 28-day sensitivity and electrical impedance spectroscopy measurement results on four similar sensors are presented. The sensitivity of the sensor is observed to be related to its double-layer capacitance and charge-transfer resistance, based on results acquired from a sensor that showed substantial sensitivity drop. Two data clusters are extracted that relate the sensor sensitivity to its impedance before and after the sensitivity drops by more than 50%.

References

18