F1000Research · 2017 · 11 citations · 8 references
Veterinary MedicineAnesthesiologyAnesthesiaAnesthetic TechniquesGeneral AnesthesiaInhalation AnesthesiaInhalational AnesthesiaAnesthetic VaporizationAnesthetic Vaporizer UseAnesthetic Wash-inFresh Gas FlowGas FlowInhalational Anesthetic AgentsMinimal Alveolar ConcentrationWash-out Dynamics
) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 - 2 - 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The "optimal FGF" for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems "optimal" taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out.
8
Static lung volumes and airway resistance reference values in healthy adults
Beate Koch, Nele Friedrich, Henry Völzke et al. · Respirology · 2012 · 35 citations