Scientific Reports · 2017 · 55 citations · 45 references
Natural SciencesBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsFluorescence ImagingFluorescence MicroscopyLight MicroscopySingle-molecule DetectionSingle-molecule ImagingSuper-resolution MicroscopyCamerasFast Image AcquisitionFluorescence Microscopy AcquisitionMolecule LocalizationSingle-molecule Imaging InstrumentationSingle-molecule Localization MicroscopySuper-resolutionSuper-resolution Fluorescence Microscopy
Many commercial as well as custom-built fluorescence microscopes use scientific-grade cameras that represent a substantial share of the instrument's cost. This holds particularly true for super-resolution localization microscopy where high demands are placed especially on the detector with respect to sensitivity, noise, and also image acquisition speed. Here, we present and carefully characterize an industry-grade CMOS camera as a cost-efficient alternative to commonly used scientific cameras. Direct experimental comparison of these two detector types shows widely similar performance for imaging by single molecule localization microscopy (SMLM). Furthermore, high image acquisition speeds are demonstrated for the CMOS detector by ultra-fast SMLM imaging.
45
Fiji: an open-source platform for biological-image analysis
Johannes Schindelin, Ignacio Arganda‐Carreras, Erwin Frise et al. · Nature Methods · 2012 · 70.9K citations
Imaging Intracellular Fluorescent Proteins at Nanometer Resolution