Technical characterization of five x-ray detectors for paediatric radiography applications

Nicholas Marshall, Maria-Helena Smet, M Hofmans, Herman Pauwels, Tom De Clercq, Hilde Bosmans

Physics in Medicine and Biology · 2017 · 14 citations · 20 references

Concepts

Abstract

; 2.5 µGy). Noise separation analysis using the polynomial model showed higher electronic noise for the DX-D35C and DRX-2530C detectors; this explains the reduced DQE seen at 0.7 µGy/image. The NIP CR detector offers notably improved DQE performance compared to the PIP CR system and a value similar to the DQE for FPD devices at the RQA3 energy.