Nonenzymatic Sensor for Lactate Detection in Human Sweat

Nikolay V. Zaryanov, E.V. Karpova, Elena E. Karyakina, Arkady A. Karyakin

Analytical Chemistry · 2017 · 148 citations · 32 references

Concepts

Abstract

High Resolution Image Download MS PowerPoint Slide For noninvasive diagnostics of hypoxia, we propose the nonenzymatic sensor based on screen-printed structures with the working surface modified in course of electropolymerization of 3-aminophenylboronic acid (3-APBA) with imprinting of lactate. Impedimetric sensor allows lactate detection in the range from 3 mM to 100 mM with the detection limit of 1.5 mM; response time is 2–3 min. Sensor sensitivity remains unchanged within 6 months of storage unpacked in dry state at a room temperature, which is unachievable for enzyme based devices. Analysis of human sweat with poly(3-APBA) based sensor is possible due to (i) much higher lactate content compared to other polyols and (ii) high sensor selectivity ( K lactate glucose < 3 × 10 –2 ). Successful detection of lactate in human sweat by means of the poly(3-APBA) based sensor has been confirmed using the highly specific reference method based on lactate oxidase enzyme (correlation coefficient r > 0.9). The attractive performance characteristics of poly(3-APBA) based enzyme-free sensors justify their future use for noninvasive clinical analysis and sports medicine.

References

32