Analysis of Two Quantitative Ultrasound Approaches

Pauline Muleki-Seya, Aiguo Han, Michael P. André, John W. Erdman, William D. O’Brien

Ultrasonic Imaging · 2017 · 17 citations · 49 references

Concepts

Abstract

There are two well-known ultrasonic approaches to extract sets of quantitative parameters: Lizzi-Feleppa (LF) parameters: slope, intercept, and midband; and quantitative ultrasound (QUS)-derived parameters: effective scatterer diameter (ESD) and effective acoustic concentration (EAC). In this study, the relation between the LF and QUS-derived parameters is studied theoretically and experimentally on ex vivo mouse livers. As expected from the theory, LF slope is correlated to ESD ([Formula: see text]), and from experimental data, LF midband is correlated to EAC ([Formula: see text]). However, LF intercept is not correlated to ESD ([Formula: see text]) nor EAC ([Formula: see text]). The unexpected correlation observed between LF slope and EAC ([Formula: see text]) results likely from the high correlation between ESD and EAC due to the inversion process. For a liver fat percentage estimation, an important potential medical application, the parameters presenting the better correlation are EAC ([Formula: see text]) and LF midband ([Formula: see text]).

References

49