Serosal Laceration During Firing of Powered Linear Stapler Is a Predictor of Staple Malformation

14

Citations

11

References

2017

Year

Abstract

We have shown that serosal laceration is a predictor of staple malformation and demonstrated the importance of precompression and slow stapling when using the powered stapling method.