Concepedia

Publication | Closed Access

Characterization of proton pencil beam scanning and passive beam using a high spatial resolution solid‐state microdosimeter

61

Citations

23

References

2017

Year

Abstract

The SOI microdosimeter with its well-defined 3D SV has applicability in characterizing proton radiation fields and can measure relevant physical parameters to model the RBE with submillimeter spatial resolution. It has been shown that for a physical dose of 1.82 Gy at the BP, the derived RBE based on the MKM model increased from 1.14 to 1.6 in the BP and its distal part. Good agreement was observed between the experimental and simulation results, confirming the potential application of SOI microdosimeter with 3D SV for quality assurance in proton therapy.

References

YearCitations

Page 1