Biomechanical Analysis of the Proximal Femoral Locking Compression Plate: Do Quality of Reduction and Screw Orientation Influence Construct Stability?

Ivan Zderic, Jong‐Keon Oh, Karl Stoffel, Christoph Sommer, Tobias Helfen, Gastón Camino-Willhuber, Geoff Richards, Sean E. Nork, Boyko Gueorguiev

Journal of Orthopaedic Trauma · 2017 · 25 citations · 24 references

DOIFull text

Open access

Abstract

Generally, malreduction led to significantly earlier construct failure. The observed failures were cut-out of the proximal screws in the femoral head, followed by either screw bending, screw loosening, or screw fracture. Proper placement of the proximal screws in anatomically reduced fractures led to significantly higher construct stability. Our data also indicate that once the screws are placed off-axis (>5 degrees), the benefit of an anatomic reduction is lost.

References

24