2017 · 12 citations · 4 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBiosensorsSensorsBiomedical InstrumentationBiomedical SensorsBiopotential MeasurementBiosensor DesignMedical InstrumentationPhysiological Signal AcquisitionAnalog Front-end CircuitsBiosensor Signal AmplificationIntegrated BiosensingMulti-modal Physiological SensingPhysiological Signal Monitoring
This paper presents a highly reconfigurable analog front-end (AFE) IC supporting multi-modal (bio)signal monitoring. By efficiently reusing core components, the reconfigurable AFE channel occupies an area of 1.1mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> while supporting four acquisition modes, i.e. biopotential (ExG), bio-impedance (BioZ), galvanic skin response (GSR) and general purpose analog (GPA). State-of-the-art sensitivity has been achieved at low power by employing both chopping and dynamic element matching (DEM). The reconfigurable AFE channel consumes 36μW maximum from a 1.2 V supply.
4
Mario Konijnenburg, Stefano Stanzione, Long Yan et al. · IEEE Journal of Solid-State Circuits · 2016 · 107 citations