Natural SciencesBiochemistryChemistryAnalytical ChemistryBiosensorsElectrochemistryBioanalytical ChemistryElectroanalytical ChemistryElectrochemical BiosensorsElectrochemical DetectionElectrochemical SensorsEnzymatic BiosensorsAmperometric BiosensorsAmperometric DetectionAmperometric SensingAmperometric SensorsCarbon Paste ElectrodesElectrochemical SensingGlucose OxidaseMethod Detection Limit
An amperometric biosensor for the indirect determination of mercury was developed by inhibition of glucose oxidase immobilized on a carbon paste electrode bulk modified with manganese dioxide 5% (m:m). Optimum conditions for biosensor were found as: operating potential 0.46 V vs Ag/AgCl, enzyme load 0.56 U/mm2 electrode surface, supporting electrolyte: 0.1 M citrate phosphate buffer (pH 7). The current response of the biosensor in hydrodynamic amperometry was linearity dependent on the concentration within 2.0 – 32.5 mg/L Hg (II) with a detection limit of 0.5 mg/L. The enzymatic activity could be fully recovered by exposure of the electrode to a solution of EDTA (0.1 M) for 2 minutes. Minor interferences were caused by Cr3+ and Zn2+. Analytical results of spiked water sample showed a good correlation with determinations by atomic absorption spectrometry where the difference is not statistically significant at 95% confidence level. The sensor yielded a stable and reproducible response more than 2 weeks when stored dry at 4 ๐C.