Quantitative Collision Cross-Sections from FTICR Linewidth Measurements: Improvements in Theory and Experiment
17
Citations
51
References
2017
Year
Ion Collision Cross-sectionsEngineeringMeasurementQuantitative Collision Cross-sectionsIon Mobility SpectrometryChemistryCrafti Cross-sectionsIon ProcessAnalytical InstrumentationCalibrationAnalytical ChemistryInstrumentationBiophysicsPhysicsBiochemistryPhysical ChemistryFticr Linewidth MeasurementsCross-sectional AreasIon MobilityRadarNatural SciencesMass SpectrometryCollision DetectionIon StructureDrug Analysis
Two corrections to the equation used in the cross-sectional areas by Fourier transform ion cyclotron resonance ("CRAFTI") technique are identified. In CRAFTI, ion collision cross-sections are obtained from the pressure-dependent ion linewidths in Fourier transform mass spectra. The effects of these corrections on the accuracy of the cross-sections obtained using the CRAFTI technique are evaluated experimentally using the 20 biogenic amino acids and several crown ether complexes with protonated alkyl monoamines. Good absolute agreement is obtained between the CRAFTI cross-sections and the corresponding cross-sections obtained using both static drift ion mobility spectrometry and computational simulations. These results indicate that the CRAFTI cross-sections obtained using the updated equation presented here are quantitatively descriptive of the size and shape of the gas-phase ions. Cross-sections that differ by less than 3% are measured for the isobaric isomers n-butylamine and tert-butylamine complexed with the crown ethers. This level of precision is similar to what has been achieved previously using traveling wave ion mobility devices. These results indicate that CRAFTI can be used to probe subtle structural differences between ions with approximately the same precision as that achieved in traveling wave ion mobility devices. Graphical Abstract ᅟ.
| Year | Citations | |
|---|---|---|
Page 1
Page 1