Publication | Open Access
Development of a silicon diode detector for skin dosimetry in radiotherapy
12
Citations
48
References
2017
Year
This work characterizes an innovative detector for in vivo and real-time skin dose measurements that is based on an epitaxial silicon diode combined with the Centre for Medical Radiation Physics (CMRP) "drop-in" packaging technology. The skin diode proved to have a water equivalent depth of measurement of 0.075 ± 0.005 mm and the ability to measure doses accurately relative to reference detectors.