High‐Performance Liquid Chromatography Method for Rich Pharmacokinetic Sampling Schemes in Translational Rat Toxicity Models With Vancomycin

MD Joshi, J. Nicholas O’Donnell, Natarajan Venkatesan, Jack Chang, Hoang Anh Nguyen, Gwendolyn Pais, RL Chapman, Brooke Griffin

Clinical and Translational Science · 2017 · 15 citations · 31 references

DOIFull text

Open access

Concepts

Abstract

A translational need exists to understand and predict vancomycin-induced kidney toxicity. We describe: (i) a vancomycin high-performance liquid chromatography (HPLC) method for rat plasma and kidney tissue homogenate; (ii) a rat pharmacokinetic (PK) study to demonstrate utility; and (iii) a catheter retention study to enable future preclinical studies. Rat plasma and pup kidney tissue homogenate were analyzed via HPLC for vancomycin concentrations ranging from 3-75 and 15.1-75.5 μg/mL, respectively, using a Kinetex Biphenyl column and gradient elution of water with 0.1% formic acid: acetonitrile (70:30 v/v). Sprague-Dawley rats (n = 10) receiving 150 mg/kg of vancomycin intraperitoneally had plasma sampled for PK. Finally, a catheter retention study was performed on polyurethane catheters to assess adsorption. Precision was <6.1% for all intra-assay and interassay HPLC measurements, with >96.3% analyte recovery. A two-compartment model fit the data well, facilitating PK exposure estimates. Finally, vancomycin was heterogeneously retained by polyurethane catheters.

References

31